Measurement time on mass number
MASS
with
mass-MODE:SAMPLE
Electrometer or Extern:
DWELL
detect-TYPE:FARAD, SEM or EXTERN
0.5, 1, 2, 5, 10, 20, 50 ms
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, 60 s
The measured value is determined by
averaging across the DWELL time.
Ion counter:
DWELL
detect-TYPE:ION-CNT
1 ms ... 60 s
Counting rate = pulse count /
DWELL
400
E-PROT
Filament protection
0.00 ... 5.00 A
When the set current threshold is exceeded the filament is switched
off and an error message is output.
125
EMI-CUR
Switch between displaying emission current and electrometer value.
OFF
Display electrometer value
ON
Display emission current
The measurement cycle must be stopped (halt).
400
EMISS
Emission current set point
OFF
Emission switched off
0.01 ... 2.00 mA
Emission current
END
With
cycle-FUNCT:CYCLE and cycle-MODE:MULTI
0 ... 63
Ending channel of the measurement cycle with
cycle-MODE:MULTI
See
TEST-CS 422, TEST-DSP
and
TEST-QMS
See
V5
See
V4
Assignment of an ion source parameter set to filament 1 or 2. Applies also to the
assignment with
SPEC±
(without filament).
FIL1, FIL2
With
ion src-MODE:NORMAL
SET 0 ... SET 3
Assignment of the ion source parameter set
Filament selection for ion sources containing two filaments
400
FILAM
Not with
config-SYSTEM-IS-TYP:AXIAL
1
Filament 1
2
Filament 2
1+2
Filament 1 in operation, filament 2 is pre-heated.
If filament 1 is defective filament 2 is automatically activated. This
results in a brief fading of the emission.
125
FILAM
filament select
switch on QME 125
→
[
3
] set to
remote
!
1
Filament 1 in normal operation / 1+2 with DEGAS
2
Filament 2
DWELL (mass)
E-PROT (emiss)
EMI-CUR (ion src)
EMISS (emiss)
END (cycle)
EPROM-T (config)
EXTRACT (v1...v6)
F-AXIS (v1...v6)
FIL1, FIL2 (ion src)
FILAM (ion src)