36
BG 805 981 BE (0201) QMG422.oi
BAUD
Only with
config-CTRL-MODE:ASCII, BIN and MODEM
300, 1200, 2400
4800, 9600, 19200
Baud rate of the RS232 interface, can always be set on the
CS 422 (300 baud only with ASCII protocol)
BEGIN
With
cycle-FUNCT:CYCLE and cycle-MODE:MULTI
0 ... 63
Start channel of the measurement cycle with
cycle-MODE:MULTI
See DIG-OUT
CALIB
Calibration factor for measured value.
±1E
-10
... ±9,99E
+10
The raw measured value is multiplied times
CALIB
Can be used for either normalizing a measured value, e.g. from 7.7 V to 10 V
(100%) or for conversion from [A] to [mbar].
In the following cases multiplication takes place only with the mantissa of
CALIB
:
•
For the computer interface with
amplif-MODE:FIX
or
detect-TYPE:EXTERN
and
mass-MODE:SCAN
or
PEAK
•
Always for the analog signals on the AO/IC 421 and
mon
→
V2
→
DIG-OUT
Deletes all pending error messages
Sets all offset values to zero and consequently disables offset correction
COPY TO CH
Copies parameters of the selected channel to another channel
0 .. 63
Target channel for copying process
SURE ?
Confirm copy function by pressing
COPY TO ALL
Copies the parameters of the selected channel to the channels
cycle-BEGIN...
-
END
.
SURE ?
Confirm copy function by pressing
COPY TO SET
Copies the IQ set activated under
ion src-FIL1 or FIL2
SET 0 .. SET 3
Target set for the copying process (only with
ion src-
MODE:NORMAL)
CP-LEV
Only with
detect-TYPE:ION-CNT
0.10 .. 1.00 V
Response threshold of the CP 400
→
17
→
under
TEST
See BAUD, MODE, NODE or SEM+FIL
CTRL
Only with
ion src-MODE:DEGAS
STOP
Switch Degas off
START
Switch Degas on
SURE ?
Confirm Degas activation with
RUN
Degas switched on
BAUD (config-CTRL)
BEGIN (cycle)
BIT (di/do-DIG-OUT)
CALIB (amplif)
CATH (v1...v6)
CLEAR (di/do-DIG-OUT)
CLEAR (error)
CLEAR (cycle-FUNCT-OFFSET)
COPY TO CH (aux)
COPY TO ALL (aux)
COPY TO SET (ion src)
CP-LEV (amplif)
CS 422 (config)
CTRL (config)
CTRL (ion src)