Alcatel Vacuum Technology France - ATH 200 User’s Manual
Draining of the water cooling circuit
Draining
Stop the pump and wait its cooling.
Then, stop the water supply.
Prepare a flexible pipe (A) with a 1/4 NPT quick female connector
(to be connected to the customer compressed air circuit) and
another flexible pipe (B) with a 1/4 NPT male connector.
Disconnect
«IN»
and
«OUT»
water quick-connect connectors on the
pump.
Connect pipe (A) at location of
«IN»
and connect the other end to
a compressed air circuit (2 to 5 bars).
Connect pipe (B) on the pump, at location
«OUT»
and place the
other end of the pipe in a vessel (total water volume will be less
than one litre).
When the water is filled out, disconnect pipes (A) and (B).
Water IN
1/4 NPT
1/4 NPT
Water OUT
A
Compressed air
Vessel
B
Rear panel
Water IN
Water OUT
Study the precautions for first use
D 00 and E 11
.
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions