2/2
D 20
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Diagnosis and trouble shooting
GB 04366 - Edition 02- Februar
y 11
INCIDENT
CAUSE
TROUBLESHOOTING
Water problem
The water temperature is
> 75°C
Î
Check the water supply.
Check that the water supply
temperature is < 25°C.
Check that the water cooling line
is not clogged or has no leaks.
)
)
)
The water flow is insufficient
Î
Check that the water supply
temperature is < 25°C.
Check that the water flow is
> 100l/h.
)
)
Other problem
Î
Call the customer service center.
)
Pump is running and yellow light indicator is lit up
INCIDENT
CAUSE
TROUBLESHOOTING
It is impossible to stop the pump
with the LOC/REM switch.
The LOC/REM switch is in «local»
mode position.
Î
Place the switch in «remote»
mode
C 10
.
)
Voltage sent between pin 1
and 2 of the J4 connector is
15 to 24 VDC
Î
Check that the voltage between
the pins 1 and 2 of the J4
connector is < 15 VDC.
)
Other problem
Î
Call the customer service center.
)
Bad vacuum or no vacuum
Foreline vacuum leak
Î
Disconnect foreline and check
the vacuum with a gauge.
Inspect the connections and
perform a helium leak test.
)
)
Other problem
Î
Call the customer service center.
)
Pump noisy
The pump is not correctly fixed
on the frame
Î
Check the fixation between the
pump and the frame.
)
The exhaust is not connected
Î
Check the connection of the
exhaust.
)
Other problem
Î
Call the customer service center.
)
Pump is running and green light indicator is lit up
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions