GB 04375 - Edition 02 Feb. 11
3/5
C 55
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Use of the RS 232 serial link
Order
Paramater
Response
Action
SEL
None
Returns the state of the parameters defined by OPT (continued)
28 :
Separator
character
30 to 43 :
Reserved
44 :
Carriage
return
•
SET
XXYZZZZ
OK
or
ERR2
Defines the internal operating parameters.
Example: #adrSETXXYZZZZ<CR>
XX =
07
: Full overhaull : 0010 to 0420 (A100L/A180L)
(0 = time x100)
Y = 0
: Alert
1
:
Alarm
ZZZZ =
Value
•
SYS
ON
or
OFF
OK
or
ERR2
Start up the pumping unit with SYSON and stop with SYSOFF.
•
STA
None
Returns the status of the internal dynamic parameters for RS232 and
RS485.
Example: #adr,
A
BCDEFGHIJ
K
,L123,M123,N12,O12,P12,Q123,R1,abcdef
g
hi
j
klm
n
opqrstu,S123
<CR>
0
:
Header
1-3
:
Address
: 000 to 9999
4
:
Separator character
5
:
ADP (J5-5 pump running)
:
0 Pump is stopped
1 Pump is running
6
:
Reserved
7 :
Reserved
8 :
Reserved
9 :
Reserved
10 :
Reserved
11 :
Reserved
12 :
Reserved
13 :
Reserved
14 :
Reserved
15
:
Management
: bit 2,1,0 = 000 = local
bit 2,1,0 = 001 = remote
(local is off) (J5-6 remote/local)
bit 4,5,6 = 0 (free)
bit 7 = 1 always
16 :
Separator
character
17-20 :
Reserved
:
0000
21 :
Separator
character
22-25 :
Reserved
:
0000
26 :
Separator
character
27-29 :
Reserved
:
000
30 :
Separator
character
•
List of controls (continued)
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions