Electricity (50/60 Hz)
A100L
A100L Energy Saving
Power supply voltage -
Low voltage
Full load current
Recommended breaker size
200 / 208 V - 3 phases - 50/60 Hz
12 A
15 A type 10000AIC
Power supply voltage -
High voltage
Full load current
Recommended breaker size
380 / 460 V - 3 phases - 50/60 Hz
7 A
15 A type 10000AIC
-
-
-
Power consumption at ultimate pressure
1.6 kW
0.8 kW
Pumping circuit
Gas inlet -
flange
material
DN 50 - ISO KF
Stainless Steel
Gas exhaust -
flange
material
DN 25 - ISO KF
Stainless Steel
Water cooling circuit
Cooling water (min.)
l/h
gpm
100
0.44
Supply pressure
(29 to 87 psig) 2-6 bars relative pressure
Min. pressure differential (inlet/outlet)
(29 psi) min 2 bar
Water temperature
10 to 25°C
Connection -
inlet type
outlet type
material
1/4 NPT female connector
1/4 NPT male connector
Stainless Steel
Water line specifications
7.5 < pH< 11 - Particle size < 0.2 mm
Solid pollution < 100mg/dm
3
Hardness < 7 milli-equivalent per liter
(350mg/l of CaCo
3
or 35°f)
Resistivity : 20 000 .cm > R > 1500 .cm
Remote interface
Remote connector 16 pins
Female contacts 16 pins
AMP 182 642-1 + AMP 182 655-1
AMP 163 084-1
Remote connector 19 pins
Female contacts 19 pins
192926-0530 ITT 192922-1340 ITT cannon
192922-1380 ITT cannon
GB 04635 - Edition 02 - Dec 11
1/2
B 01
Hook-up requirements
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions