Alcatel Vacuum Technology France - ATH 200 User’s Manual
Remote control plug connection
(dry contacts / 19 pin plug)
Study the preliminary precautions
B 00
.
•
Dry contact outputs are rated 50 V DC - 1 A max.
Over voltage and over current can cause internal electrical damages.
Monitor remote control plug voltage below 50 V and current below 1 A.
When units containing control circuits are equipped with dry contact outputs,
it is the responsability of the customer to use these outputs in compliance
with extra low voltage installation and safety standards (SELV).
The remote control function allows:
Control of pumping functions START/STOP and variable rotation
speed.
Remote monitoring of pump status through auxiliary dry contacts
(50 V DC - 1 A).
These contacts can be used to control automatic functions.
Monitor pump warnings and hazards on the remote control plug.
These contacts can be monitored or used by a system external to the
pump.
Remote control
connector location
The remote control connection is done via the «remote control connector»
located at the rear panel of the pump.
Plug type:
192 926-0530 ITT 192 922-1340 ITT cannon
+ female contacts 19 pins 192 922-1380 ITT cannon
Remote control connector
p
Functions
Make sure that the main switch is off to avoid any short circuit during remote
control plug disconnection/connection.