Alcatel Vacuum Technology France - ATH 200 User’s Manual
Electrical connection
GB 04358 - Edition 03 - Dec. 11
Female plug connection
Connect the main power female plug on the pump main power male
connector and lock it as follows:
Screw the ferrule.
Fit the pipe clamp on the female connector collar (position the fixing
holes correctly).
Secure the two CS M3 screws.
Female plug wiring
Connect the main cable (customer supply) to the female main power
plug using wires with following specifications:
Use AWG-14 (or 2,08 mm
2
) wires,
Connect the green lead to the ground on the plug,
Connect the other 3 leads to the 3 phases (U, V and W).
The pump will automatically correct with autophase correction
fixture.
Main power plug at rear panel
Remove bracket before disconnecting
and install it after connection
Main power
plug
Ferrule
Bracket
CS M3 screw
Low voltage electrical
connection
Usual voltage
200/208 V 50/60 Hz
Low voltage
Max. current
12 A
Max. current measured on the pump.
The pump is supplied with an equipped main power cable separately
packaged
B 10
.This last must be connected on the main power line
according to the following diagram:
1
2
3
4
PE
earth
W
V
U
not used