GB 04363 - Edition 03 - Dec 11
1/1
C 20
Alcatel Vacuum Technology France - ATH 200 User’s Manual
Starting mode definition «Autostart»
Study the preliminary precautions
B 00
.
•
Pump start up
The pump is set at factory as follows:
Control mode
Starting mode
Local
Autostart
Configure the ‘LOC/REM’ switch in «Local control mode».
Put the main switch on position
1
.
The following operations will perform automatically:
Pump starts.
Sensor processing.
Pumping shut down
Put the main switch on position
0
.
The following operation will perform automatically:
Pump stops.
Note : When the stop is caused by an hazard, correct the fault, put the main switch on
position 0, then 1 to restart the pump.
Do not move a pump when it is in operation.
Operating mode,
§ selection of the pump control mode
C 10
The performance and operational safety of this product are guaranteed
provided it is used normally in the operating conditions defined in this
operating instruction manual.
It is the customer’s task to:
- train operators to use the product if they do not speak the language the
manual is written in,
- ensure operators know the safe practices to apply when using the
product.
At the initial start-up, provide protection against the risk of crushing related
to the rotating parts in the inlet.
A non-powered pump can be driven by another pump in rotation (risk of
crushing). Always connect the pump inlet port before starting the pump.
The pump is tested and filled with a synthetic oil charged at factory.
Do not modify the oil level.
Auto-restart.
It is the responsibility of the user to take all the actions required to prevent
risks resulting from this type of operation. The user must provide a device
(integrated in the equipment/host tool) to warn or to avoid this restart.
adixen Vacuum Products - A100L - Dry Pump - Operating Instructions