Installation . 21
Cell Holder Installation
PTP-1
The PTP-1 consists of a single sample cell holder which is installed in the sample beam path
in the spectrometer.
Before installing the single sample cell holder, remove any accessory currently installed in the
sample compartment.
1.
Remove the standard sample cell holder by simply screwing out the mounting posts.
2.
Lever out the cap of the plastic plate with a screwdriver (Figure 2).
3.
Feed the connecting cable through the hole.
4.
Fix one end of the connecting cable to the bottom of the cell holder (Figure 3) and the
other end to the rear of the control unit (Figure 6).
Figure 2 Cap covering hole for connecting cable (Lambda 25)
Cap