22 . Peltier Temperature Programmer User’s Guide
Figure 3 Bottom of PTP-1 cell holder with connecting cable attached (Lambda 25)
PTP-1+1
The PTP-1+1 consists of a single sample cell holder and a single reference cell holder. The
single sample cell holder is installed in the sample beam path in the spectrometer, the single
reference cell holder is installed in the reference beam path in the spectrometer.
Before installing the cell holders, remove any accessory currently installed in the sample
compartment.
1.
Remove the standard sample cell holder and the standard reference cell holder by
simply screwing out the mounting posts.
2.
Lever out both caps on the plastic plate with a screwdriver (Figure 2).
3.
Push both connecting cables under the spectrometer and then feed one cable through
the hole in the sample beam path, and one cable through the hole in the reference
beam path.
4.
Fix one end of a connecting cable to the bottom of the sample cell holder (Figure 3) and
the other end to the
Sample
socket on the rear of the control unit (Figure 4).
5.
Fix one end of a connecting cable to the bottom of the reference cell holder and the
other end to the
Reference
socket on the rear of the control unit (Figure 4).
Summary of Contents for PTP-1
Page 1: ...Peltier Temperature Programmer User s Guide...
Page 5: ...Introduction...
Page 10: ...10 Peltier Temperature Programmer User s Guide...
Page 11: ...Warnings and Safety Information...
Page 18: ...18 Peltier Temperature Programmer User s Guide...
Page 19: ...Installation...
Page 30: ...30 Peltier Temperature Programmer User s Guide...
Page 31: ...Description...
Page 36: ...36 Peltier Temperature Programmer User s Guide...
Page 37: ...Operation...
Page 47: ...Configuration and Maintenance...