Installation . 27
System Connections
Figure 7 System component connections
Connecting a Flowing Water Source
CAUTION
A minimum water flow of 200 ml/min is mandatory for proper
operation.
1.
Cut the PVC
tube provided into the required number of pieces.
For the PTP-1 and PTP-6 you will need two equal lengths of tubing. For the PTP-1+1
and PTP-6+6 you will need 4 equal lengths of tubing.
2.
Pass the tubing through the front of the sample compartment and connect it securely to
the olives of the heat sinks.
Ensure that the tubing does not obstruct the light path.
3.
Connect the other ends of the tubing to the water source and drain.
PTP
Spectrometer
PC
Cell Holder(s)
RS232 Serial Cable
Cell Holder Cable(s)