Using the Spectrometer . 57
The Single-Cell Holder
Description
There are two single-cell holders provided with the instrument (see Figure 25): one for the
sample beam and one for the reference beam. The single-cell holders are mounted on a
plinth to bring them in line with the radiation beam.
Figure 25 Single-cell holder (B0505071)
Installing the Single-Cell Holder
To install the plinth in the sample compartment:
1.
Lower the plinth so that the two locating holes slip onto the two locating pins on the
baseplate in the sample compartment, as shown in Figure 26.
The plinth fits one way round only.
Figure 26 Fitting the plinth to the baseplate in the sample compartment
2.
Screw the two thumbscrews by hand into the baseplate.
Locating pin
Thumbscrew
Thumbscrew
Locating pin
Locking screw
for horizontal
alignment
Vertical-adjustment
screw
Milled post
Lifter
Locking screw for
horizontal alignment
Summary of Contents for Lambda 1050
Page 1: ...HIGH PERFORMANCE LAMBDA SPECTROMETERS Hardware Guide UV VIS SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 11: ...Safety Practices ...
Page 25: ...System Description ...
Page 32: ...32 High Performance Lambda Spectrometers Hardware Guide ...
Page 33: ...Installing your Lambda Spectrometer ...
Page 54: ...54 High Performance Lambda Spectrometers Hardware Guide ...
Page 55: ...Using the Spectrometer ...
Page 64: ...64 High Performance Lambda Spectrometers Hardware Guide ...
Page 65: ...Maintenance ...
Page 76: ...76 High Performance Lambda Spectrometers Hardware Guide ...
Page 77: ...Appendices ...
Page 82: ...82 High Performance Lambda Spectrometers Hardware Guide ...