46 . High-Performance Lambda Spectrometers Hardware Guide
Slits:
There are no user-defined settings in the Slits calibration.
Zero Percent T:
There are no user-defined settings in the Zero Percent T calibration.
5.
Click
Calibrate
.
The calibrations are performed in the order listed in the table. Messages below the table
show the progress of the calibration.
If you are performing the Zero Percent T calibration on a Lambda 650/750, you will be
prompted to place the magnetic beam blocks supplied over the ports in the sample
compartment wall.
When a calibration has passed, a green check mark is displayed in the first column of
the table. If the test fails, a red cross is displayed. In such an event please contact your
local PerkinElmer Service Engineer.
6.
When the calibration has finished, click
Close
to close the dialog.
The date and time of the calibration are recorded in the instrument event log.
NOTE: It is not possible to print the calibration results.
Your instrument is now ready to be used.
Summary of Contents for Lambda 1050
Page 1: ...HIGH PERFORMANCE LAMBDA SPECTROMETERS Hardware Guide UV VIS SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 11: ...Safety Practices ...
Page 25: ...System Description ...
Page 32: ...32 High Performance Lambda Spectrometers Hardware Guide ...
Page 33: ...Installing your Lambda Spectrometer ...
Page 54: ...54 High Performance Lambda Spectrometers Hardware Guide ...
Page 55: ...Using the Spectrometer ...
Page 64: ...64 High Performance Lambda Spectrometers Hardware Guide ...
Page 65: ...Maintenance ...
Page 76: ...76 High Performance Lambda Spectrometers Hardware Guide ...
Page 77: ...Appendices ...
Page 82: ...82 High Performance Lambda Spectrometers Hardware Guide ...