System Description . 29
The radiation passing alternately through the sample and reference beams is reflected by
mirrors M11, M12, M13, and M11′, M12′, M13′, respectively of the optics in the detector
assembly onto the appropriate detector. Mirror M14 is rotated to select the required detector.
A photomultiplier tube (PMT) detector is used in the UV/Vis range, while a lead sulfide (PbS)
detector is used in the NIR range. Detector change is automatic during monochromator
slewing.
NOTE: The PbS detector is not available in the Lambda 650 or 850 spectrometers.
At the cell plane, each radiation beam is approximately 12 mm high. The width of the
radiation beams is dependent on the slit width. At a slit width of 5 nm each radiation beam is
approximately 4.5 mm wide. To permit minimum sample volumes to be measured in micro
cells, the height of the radiation beam must be reduced in the active cell area.
A common beam mask (CBM) is mounted between the slit assembly (SA) and mirror M7. This
mask restricts the cross-section of both the sample beam and the reference beam in the
respective cell area. The radiation beam can be reduced from the maximum height of
11.7 mm to 0.0 mm in 50 steps.
During all scanning operations, the monochromators stop slewing while a filter, source, or
detector change is in progress.
NOTE: You can select the source or detector change wavelength within a defined
wavelength range. This facility is useful if a feature of special spectral interest is
located at one of the default change wavelengths.
The spectrometer scans from higher toward lower wavelengths.
There is an optional depolarizing filter (DP) accessory, which can be swung into the beam.
There are automatic internal attenuators for the Lambda 850 and 950 spectrometers that can
be swung individually into the sample beam (SBA) and the reference beam (RBA). The
attenuators have the values: 0%, 1%, 10% and 100%. The Lambda 650 and 750
spectrometers require manual attenuators to be placed over the ports in the sample
compartment wall as needed (L6020277 – 10% attenuator, L6020276 – 1% attenuator).
Lambda 1050 Spectrometer
The optical system of the Lambda 1050 is the same as for the Lambda 950 spectrometer,
with three exceptions.
First, there is a source doubling mirror M0 in the lamp compartment, designed to increase
the intensity of the light from the halogen lamp (HL) that is reflected by mirror M1 to mirror
M2.
Summary of Contents for Lambda 1050
Page 1: ...HIGH PERFORMANCE LAMBDA SPECTROMETERS Hardware Guide UV VIS SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 11: ...Safety Practices ...
Page 25: ...System Description ...
Page 32: ...32 High Performance Lambda Spectrometers Hardware Guide ...
Page 33: ...Installing your Lambda Spectrometer ...
Page 54: ...54 High Performance Lambda Spectrometers Hardware Guide ...
Page 55: ...Using the Spectrometer ...
Page 64: ...64 High Performance Lambda Spectrometers Hardware Guide ...
Page 65: ...Maintenance ...
Page 76: ...76 High Performance Lambda Spectrometers Hardware Guide ...
Page 77: ...Appendices ...
Page 82: ...82 High Performance Lambda Spectrometers Hardware Guide ...