Installing your Lambda Spectrometer . 53
Figure 22 Sipper electrical connection
Pin Configuration
There are two 15-pin connectors fitted to the connector panel of the high-performance
Lambda spectrometer for the connection of accessories. The connectors are identically
configured so that the accessories can be connected to any of them. The pin numbering is
shown in Figure 23, and the configuration is given in the following table.
Figure 23 Pin configuration of 15-pin connector
Pin Configuration Description
Pin Configuration
Description
1
GND D
Ground (Digital)
logic
9
Not used
2
+5 V
Logic power
supply
10 Not used
3
Not used
11 +12 V
Analog power
supply
4
GND P
Ground Power
12 GND A
Ground Analog
5
+24 V
Power Supply
13 -12 V
Analog power
supply
6
Not used
14 SCL
Serial clock
7
Not used
15 SDA
Serial data
8
Not used
Sipper
Spectrometer
connector panel
Summary of Contents for Lambda 1050
Page 1: ...HIGH PERFORMANCE LAMBDA SPECTROMETERS Hardware Guide UV VIS SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 11: ...Safety Practices ...
Page 25: ...System Description ...
Page 32: ...32 High Performance Lambda Spectrometers Hardware Guide ...
Page 33: ...Installing your Lambda Spectrometer ...
Page 54: ...54 High Performance Lambda Spectrometers Hardware Guide ...
Page 55: ...Using the Spectrometer ...
Page 64: ...64 High Performance Lambda Spectrometers Hardware Guide ...
Page 65: ...Maintenance ...
Page 76: ...76 High Performance Lambda Spectrometers Hardware Guide ...
Page 77: ...Appendices ...
Page 82: ...82 High Performance Lambda Spectrometers Hardware Guide ...