Initialization errors
If your instrument generates one of the following error messages, switch off the instrument,
then switch the instrument on again. Consult your PerkinElmer service representative if the
problem persists.
Error message
Action
Program Integrity Failed
The instrument’s firmware may be corrupt, so must be
reinstalled.
Instrument Configuration Failed
Suggests that one or more internal devices is failing. If
the problem persists, note any associated error
messages and consult your PerkinElmer service
representative.
Scan Failed
If the problem persists, note any associated error
messages and consult your PerkinElmer service
representative.
Laser Failed
Replace Laser
Consult your PerkinElmer service representative.
The beamsplitters may be missing, the laser fringe pre-
amplifier may have failed, or the laser itself may have
failed.
Refer to
Replacing the Laser and Power Supply
page 91.
Source Failed
Replace Source
on page 83.
Calibration Failed
Too little energy may be reaching the detector. Make
sure the sample compartment is clear.
Summary of Contents for Frontier IR
Page 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Frontier IR Dual Range Systems User s Guide...
Page 11: ...Warnings and Safety Information...
Page 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Page 34: ...34 Frontier IR Dual Range Systems User s Guide...
Page 35: ...Unpacking and Installation...
Page 45: ...Using the Spectrometer with Spectrum...
Page 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Page 57: ...Routine Maintenance...
Page 78: ...78 Frontier IR Dual Range Systems User s Guide...
Page 79: ...Advanced Maintenance...
Page 94: ...94 Frontier IR Dual Range Systems User s Guide...
Page 95: ...Appendices...