66 . Frontier IR Dual-Range Systems User's Guide
Renewing the instrument desiccant
CAUTION
Old, used desiccant releases moisture and can cause catastrophic failure
of KBr optics.
Do not use damaged packs of desiccant. Make sure that the packs you
use have not been left in contact with the air.
You can purchase disposable desiccant kits from PerkinElmer (part number N0171159). A kit
contains two packs of desiccant, and three kits are required.
1.
Inspect the humidity indicator card in the plastic bags in which the spare desiccant
packs are packed. If the card indicates humidity in the bag, discard the desiccant pack.
2.
Remove the sample area cover, if fitted, by opening the cover, pressing the clip and
pulling the cover vertically to remove it (Figure 30).
Figure 30 Removing the sample compartment cover
3.
Undo the two captive screws securing the desiccant cover (Figure 31).
Figure 31 Captive screws securing the desiccant cover
Screw
Screw
Sample cover
release clip
Summary of Contents for Frontier IR
Page 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Frontier IR Dual Range Systems User s Guide...
Page 11: ...Warnings and Safety Information...
Page 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Page 34: ...34 Frontier IR Dual Range Systems User s Guide...
Page 35: ...Unpacking and Installation...
Page 45: ...Using the Spectrometer with Spectrum...
Page 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Page 57: ...Routine Maintenance...
Page 78: ...78 Frontier IR Dual Range Systems User s Guide...
Page 79: ...Advanced Maintenance...
Page 94: ...94 Frontier IR Dual Range Systems User s Guide...
Page 95: ...Appendices...