Advanced Maintenance . 87
Installing Filters in the Filter Wheel
The following procedure describes how to install a filter in the filter wheel. The same
procedure should be followed when replacing a filter or the methane cell.
The filter wheel assembly is removed from its mounting and placed on the standard sample
slide baseplate. It therefore may be necessary to remove the sample slide.
WARNING
Switch off the mains power supply to the spectrometer, wait 60 seconds,
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
1.
Set the
Filter Wheel
to Position 1 on the Setup Instrument BeamPath tab in Spectrum
software.
Position 1 will be labelled as
None 15000 – 0 cm
-1
and be the first option in the
drop-down list.
2.
Open the main cover of the instrument as detailed on page 80 and identify the filter
wheel (Figure 51).
Figure 51 Filter wheel location
Filter wheel
Filter wheel
Summary of Contents for Frontier IR
Page 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Frontier IR Dual Range Systems User s Guide...
Page 11: ...Warnings and Safety Information...
Page 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Page 34: ...34 Frontier IR Dual Range Systems User s Guide...
Page 35: ...Unpacking and Installation...
Page 45: ...Using the Spectrometer with Spectrum...
Page 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Page 57: ...Routine Maintenance...
Page 78: ...78 Frontier IR Dual Range Systems User s Guide...
Page 79: ...Advanced Maintenance...
Page 94: ...94 Frontier IR Dual Range Systems User s Guide...
Page 95: ...Appendices...