90 . Frontier IR Dual-Range Systems User's Guide
9.
Refit the filter wheel to its spindle and secure it with the set screw.
CAUTION
Make sure that you tighten the set screw onto the flat of the shaft.
If the set screw is tightened onto a round portion of the shaft the filter
wheel may loosen.
10.
Refit the sample holder onto the baseplate with the knurled screw.
11.
Close and secure the main cover.
Refer to steps 11 to 15 on page 82.
12.
Select the filter you would like to use in the beam path from the
Filter Wheel
drop-down list on the Setup Instrument BeamPath tab in Spectrum software.
NOTE: If you wish to run a background scan with a filter in place, you must set the
Scan
Type
to
Background
on the Setup Instrument Basic tab, rather than using the
Background
toolbar button
. See the Spectrum on-screen Help for more
information.
Summary of Contents for Frontier IR
Page 1: ...FRONTIER IR DUAL RANGE SYSTEMS User s Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 Frontier IR Dual Range Systems User s Guide...
Page 11: ...Warnings and Safety Information...
Page 23: ...An Overview of the Frontier IR Dual Range Spectrometers...
Page 34: ...34 Frontier IR Dual Range Systems User s Guide...
Page 35: ...Unpacking and Installation...
Page 45: ...Using the Spectrometer with Spectrum...
Page 52: ...52 Frontier IR Dual Range Systems User s Guide Figure 21 Atmospheric suppression...
Page 57: ...Routine Maintenance...
Page 78: ...78 Frontier IR Dual Range Systems User s Guide...
Page 79: ...Advanced Maintenance...
Page 94: ...94 Frontier IR Dual Range Systems User s Guide...
Page 95: ...Appendices...