Advanced Maintenance . 91
Replacing the Laser and Power Supply
WARNING
Switch off the main power supply to the spectrometer, wait 60 seconds
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
WARNING
Do not attempt to separate the Laser from the power supply. The Laser
retains an electrical charge even when the main power is switched off.
1.
Open the main cover of the instrument as detailed on page 80 and identify the laser
and the laser power supply.
Figure 56 Laser and power supply
Laser
Power supply
Laser
Power supply