CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
30
PERISO SA
operate the device for about 10-15
minutes and then verify that the
light is green/blue again; again off
the device; reposition the handpiece
on the arm; reboot the device and try
that the flow of the magnetic field is
correct.
The device turns on, but the handpiece does
not emit radio frequencies.
The hand piece is properly
connected to the power module?
The connection cable is connected
correctly to the power
module/controller?
Are the electrodes, neutral plates
and the cables of the neutral plates
intact? Verify its integrity by
performing the tests with the
various electrodes or the second
return plate.
Check the correct amount of water in
the radiator, making sure that the
indicator light is green/blue color.
The red light indicates missing
water or below the necessary level.
If the light is red: Turn off the device
by unplugging it from the mains;
Open the tank cap and top up the
water in adequate quantities, using
the filler special tool provided with the
device; close the cap; place the
dispenser handle lying on the
ground; restart the device by
pressing start on the display; to
operate the device for about 10-15
minutes and then verify that the
light is green/blue again; again off
the device; reposition the handpiece
on the arm; reboot the device and try
that the provision of frequencies is
correct.
The device turns on, but the handpiece does
not emit.
Are the cables between the
controller and the power module
correctly connected?
Is the handpiece and its connection