CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
29
PERISO SA
Make sure there are no devices that
emit strong electromagnetic fields
nearby.
Turn off the device, disconnect all
connection cables, reinsert them,
switch the device back on.
On the display of the horizontal or vertical
lines are shown that make partial reading of
the information.
Is the cable between the device
controller and the power module
connected properly?
Check the display of the same
functionality in different points.
Ensure that the display is not greasy
or dirty and that there is no cover
film.
Emission
The device turns on, but the handpiece
does not supply the magnetic field output.
Has a work program been set and
has been pressed START on the
display?
The handpiece is properly
connected to the power module?
Check the integrity of the cable.
Check the integrity of the
connector.
The connection cable is connected
correctly to the power
module/controller?
Check the correct amount of water in
the radiator, making sure that the
indicator light is green/blue color.
The red light indicates missing
water or below the necessary level.
If the light is red: turn off the device
by unplugging it from the mains;
open the tank cap and top up the
water in adequate quantities, using
the filler special tool provided with the
device; close the cap; place the
dispenser handle lying on the
ground; restart the device by
pressing start on the display; to