6.3 Troubleshooting
Problem
Potential Solution
Power source
The device does not switch on.
Is the power cable connected to
230V 50 / 60Hz?
Has the key been turned to the ON
position?
Check that the emergency button it
is not pressed.
The display does not switch on.
Has the key been turned to the ON
position?
The green LED, on the back of the
power module, are turned on but
nothing is shown on the display?
Make sure the cable connections
between the device controller and
the power module are correct and
Type of Verification
Frequency
Connection
to the mains
supply
Power on
Integrity of neutral plates cables
Daily
Daily
Yes
No
Integrity of the handpiece cables
Integrity of the power supply cable
Daily
Daily
No
No
Integrity of power connection cable – controller module
Monthly
No
Integrity of neutral plates
Each use
No
Integrity of the handpiece
Each use
No
Integrity of the electrodes
Each use
No
Level of water in the tank
Each use
No