CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
25
PERISO SA
6. MAINTENANCE AND TROUBLE SHOOTING
6.1 Cleaning and disinfection
The cleaning of the device’s frame must be done with a damp cloth using neutral and
non-aggressive detergents.
Do not use alcohol for cleaning.
The electrodes and the neutral plates should be cleaned and disinfected with a damp
cloth, warm water and a non-aggressive detergent after each use in order to prevent any
deposits from solidifying and difficult to remove as well as potentially damage the
accessories and the handpiece.
After cleaning the electrodes and the neutral plates make sure to dry them completely.
Clean the handpiece, the electrodes and the neutral plates on first use.
Make sure that the device is disconnected from the power supply when cleaning
it.
Before cleaning the neutral plates, disconnect them from the device.
Only use the cream produced by Periso SA for treatments with the CTU Mega 20
Diamagnetic Pump.
6.2 Guide to the correct maintenance
The device must be subjected to annual review and must be carried out by the
Periso manufacturer or by qualified personnel from Periso.
The device must be installed and put into service in accordance with the
information contained in this manual.
Make sure that it is disconnected from the mains during maintenance of the
device.
Do not leave the equipment exposed to humidity, rain, direct sunlight.
Do not spill liquids on the equipment and device and do not place any objects on
the device. Do not leave voltage cables in tension in points of passage, to prevent
accidental damages.
Connect the device only to a system that complies with national standards of the
user’s Country.
In case of water flow interruption the pump stops and the magnetic field is
interrupted. A beep sounds and a red light comes on at the back of the device.
Check with each use that the red light on the back of the power module is not on (insufficient
radiator water level) and that blue / green lights are on. If the red light is on, remove the red cap
and fill the tank using only demineralised water and the supplied bottle.