CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
5
PERISO SA
1.3 Safety precautions and warnings
Before connecting the CTU Mega 20 Diamagnetic Pump device to the power
supply you must ensure that it has the following characteristics:
Mains voltage 230 Volt (+/- 10%)
Mains frequency 50-60 Hz
Network Consumption 2200 VA (800 W) 9,5 A
It is also necessary that the surrounding conditions are the following:
Room temperature from 10 °C to 30 °C
Relative humidity from 30% to 75% non-condensing
Atmospheric pressure from 700 to 1032 hPa
The environmental storage conditions must instead be the following:
Room temperature from 0 °C to 40 °C
Relative humidity from 10% to 80%
It should be avoided the installation of the equipment in the following cases:
Near heat sources
Exposed to rain or humidity
Exposed to direct sunlight
The Manufacturer accepts no responsibility for damage or malfunction resulting
from:
Supply mains do not conform to the above
Unsuitable environmental conditions
Do not use the device:
In the presence of combustible gases, flammable vapors, in rooms with
oxygen or detonating atmosphere
In the presence of flammable anesthetic mixture with air, oxygen or nitrous
oxide
Near equipment with strong electromagnetic emissions or sensitive to
electromagnetic emissions
Near to or over other equipment
Keep the device away from splashing water, even accidental.