CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
23
PERISO SA
Hypotension
: For people with low blood pressure (hypotension): the arterial pressure should be
checked before and after treatment and, if necessary, the duration of the session should be
decreased
and
the
time
between
two
sessions
should
be
increased.
The majority of the contraindications listed above are based on a series of non-
existing official results but only on hypothetic or potentially of patients treated with
these conditions. In any case, in the presence uncertainty or of undefined symptoms,
if in doubt, the patient should consult the doctor.
5.3 Precautions
Begin treatment only if the patient has no ferromagnetic deposit/objects or similar
materials in the area to be treated and focus on knowing the health condition of the
subject to be treated.
Ensure that both the operator and the patient do not wear any metallic object.
The device must be operated only by personnel trained and qualified to use the
device.
In case of absence of personnel trained in the use of the machine, turn off the
device by the general switch on the system.
Handle with care the CTU Mega 20 handpiece, making sure not to cause damage
(scratches, drops, bumps, etc.).
After each use, please check the integrity of the connection cables and associated
connectors, and all moving or fixed parts to assure good working order.
In case of damage or deterioration of the cables or power supply, contact support.
The use of accessories, transducers and cables other than those supplied by the
manufacturer, is prohibited. It could result in increased or decreased emissions or
other compromising the optimal operating/performance of the device.
Set best working temperature level.
Make sure that the CTU Mega 20 has no contact with liquids.
In the event of malfunctions, stop the treatment and notify the technical service.
In the case of smoke emission, excessive heat or noise from the device it must be
stopped immediately.
In case of irregular temperature increase of the operating handpiece, you must
stop the treatment immediately, since this is an index of malfunction.
You must not use the device in presence of fuel gases, flammable steams, in
room with oxygen or detonating atmosphere.
In case of falling from a height exceeding 30 cm of the insulated capacitive
electrode, avoid the further use of this electrode.
The patient must not come into contact with earthed metal parts during treatment,
or with an appreciable ground capacity.
A low level of dispensing, or incorrect operation, can be caused by a faulty
application of the neutral plate. In this case, check the connections.
The power of electrodes can change during use.