CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
19
PERISO SA
Neutral plates
Depending on the area of the body to be treated, you can select "single plate" or
"double plate".
Radiofrequency power
The power adjustment from 0% to 100%, favors biostimulation processes and re-
energizing of tissues.
The procedure for the correct adjustment of the power is as follows:
1) Select the area of the body being treated. Prepare the patient in the most
appropriate position.
2) Place the neutral plate in position perpendicular to the CTU Mega 20
handpiece. Apply on neutral plate the right amount of the specific cream, such
as to entirely cover the surface of the plate.
3) Choose the capacitive or resistive electrode to be mounted at the tip to the CTU
Mega 20 handpiece, depending on the region and the pathology to be treated.
4) Select the emission frequency HF (High Frequency) or LF (Low Frequency).
5) Spread on the area to be treated an adequate amount of proper cream, to
facilitate the sliding of the CTU Mega 20 handpiece during treatment.
6) Place the CTU Mega 20 handpiece on the treatment area and slowly raise the
power by using the gray power meter on the front of the display.
It is important to remember that with the CTU Mega 20 there is the automatic recognition
of the capacitive and/or resistive electrode, so it must not be preselected
on the display.