33
1 = yes) turns on count mode (0 = no, 1 =
yes).
CLS - CaLibrate Secondary
COMMAND:
CLS 0.002 2
Sets secondary weighing res. to .002KG and
units to KG.
CLE - CaLibration End
COMMAND:
CLE <ENT>
Ends the calibration process and stores the
results in the unit's internal memory. The
unit will then self-test and return to normal
operation.
RESPONSE TO TERMINAL:
"Saving CAL Data If data acceptable.
CAL Completed"
CALIBRATION EXAMPLE
Using the example commands from the previous page, a typical sequence of calibration
commands might be:
1. CLP 10.0 0.002 5.0 1 Enters the primary calibration factors
2. CLU 0
Selects print while Unstable for non-NTEP
3. CLW 3
Enters the weight calibration mode
4. CLM 0 1 1
Enters no Canadian, auto zero, count yes
5. CLS 2 0.001 Enters the secondary calibration factors
6. CLE
Ends the calibration process, stores results
See the individual command descriptions for complete command and response information.