24
8
REFERENCE INFORMATION
Product name
Alternating pressure air flotation mattress
Model No.
KE-918AW
Type of protection for electrical shock
Class I device
Level of protection for electrical shock
Type BF applied part
Protection code
IEC 60529 IP21
Mattress
Size (cm)
Width x Length x Thick-
ness
86 x 200 x 17
Product weight (kg)
8.5
Materials
Top cover
Polyester 70%, polyurethane 30% (antibacterial, flame-resistant, waterproof, and moisture-permeable)
Glide sheet material
Polyester 100%
Top cushion
Polyurethane (flame-resistant)
Bottom cover
Polyester 90%, polyurethane 10% (antibacterial and flame-resistant)
Pump cover
Polyester 90%, polyurethane 10% (antibacterial, flame-resistant, and waterproof)
Sensor cover
Polyester 70%, polyurethane 30%
Bottom cushion/
foot cushion
Polyurethane (flame-resistant)
Tube
PVC 100% (antibacterial and flame-resistant)
Air cell
Polyurethane 100% (antibacterial and flame-resistant)
Coupling
PC/PBT resin (flame-resistant)
Sensor case
PC/PBT resin (flame-resistant)
Sensor sheet
PET resin
Pump
Product weight (kg)
3.4
Case materials
ABS resin (flame-resistant and chemical-resistant)
Bed link
cable
Product weight (kg)
0.1
Length (cm)
65
Materials
PVC resin
Operation
panel
(accessory)
Product code
RC-2030W
Protection code
IEC 60529 IPx4
Size (cm)
Width x Length x
Thickness
5.6 x 5.6 x 17 (including hook)
Product weight (kg)
0.2
Cable length (cm)
200
Materials
Case
ASA resin
Hook
POM resin
Surface sheet
PC resin
Cable
PVC resin
Operating
environment
conditions
Temperature/humidity/
Atmospheric pressure
0˚C to 40˚C / 30% to 85%RH / 620 hPa to 1060 hPa
Storage
environment
conditions
Temperature/humidity/
Atmospheric pressure
Mattress/Pump/Bed link cable
-10˚C to 60˚C / 0% to 85%RH / 620 hPa to 1060 hPa
Operation panel (accessory)
-10˚C to 50˚C / 30% to 90%RH / 620 hPa to 1060 hPa
User’s weight (kg)
20 to 180
Power supply
DC port: 24 VDC, 1.65 A / bed link port: 30 VDC, 1.32 A
Service life
5 years (based on Paramount Bed data), excluding consumable items
Essential performance:
Adjust and maintain pressure in air cells
Frequently used functions: Alternating pressure and changing position function
Classification and Standards
The manual hospital bed is designed and manufactured according to the following equipment classification and standards.
•
Safety standards applied : IEC 60601-1:2005, AMD1:2012