1
2
4
9
6
8
3
5
7
1
ALTERNATING PRESSURE button
Sets alternating pressure.
is displayed in the AIR MAT-
TRESS home screen while alternating pressure is set.
2
DEHUMIDIFY button
Sets dehumidify.
is displayed in the AIR MATTRESS
home screen while dehumidify is set.
3
BACK SUPPORT (BS) button
Sets the back support function.
is displayed in the AIR
MATTRESS home screen while the back support function is
set.
4
SMALL FLOW (SF) button
Switches the setting between small flow and inflate.
is
displayed in the AIR MATTRESS home screen while small
flow is set.
5
SF SENSOR OFF MODE button
Sets SF sensor off mode.
is displayed in the AIR MAT-
TRESS home screen while SF sensor off mode is set.
6
AUTO MAX INFLATE SETTING button
Sets auto max. inflate mode.
is displayed in the AIR
MATTRESS home screen while auto max. inflate mode is
set.
7
AUTO button
Sets the automatic softness control function.
is displayed
in the AIR MATTRESS home screen while automatic soft-
ness control is set.
8
MANUAL button
Sets the manual softness control function.
is displayed
in the AIR MATTRESS home screen while manual softness
control is set.
9
FILTER CHECK button
The button is displayed when the filter operating time ex
-
ceeded the recommended replacement time (8,800 hours).
Press this button to reset the filter operating time.
* For details on replacing the filter, see page 19.
* For the details on the filter for cococia, contact Paramount
Bed (see page 31) or your distributor.
1
Fold back the sheet in the vicinity of the pump.
2
Open the CPR zipper.
3
Pull out the couplings (upper and lower) toward you while
pressing the CPR levers.
Couplings (upper and lower)
CPR lever
Coupling
Sheet
CPR zipper
1
3
2
4. How to use the CPR function
Air can be discharged from the air mattress quickly by using the CPR function in case of emergency, and cardiac massage can be started in 30 seconds.
17
Function settings
Touch the
button in the AIR MATTRESS home screen to display the air mattress settings screen.
4