20
2. Inspection
Filter check mode
This mode allows you to check and reset the filter operating time.
1
Press and hold down the Setting button.
2
The “FIL” indication and filter operating time indication are
displayed alternately in the status display screen.
3
Press the Setting button. The filter check mode will also end
after 1 minute elapses.
[How to reset the filter operating time]
Press and hold down the Max. inflate mode button for 3 sec
-
onds in the filter check mode. A beep sound is output and the
filter operating time indication becomes “0.”
Note
!
z
The filter operating time can also be reset using
the BED Navi (see page 17) .
*
The filter operation time is indicated starting from the 100th digit.
Examples: 160 hours is indicated as “1” and 80 hours is indicated as “0”
*
When the filter is used continuously, the approximate timing for re
-
placing the filter is when “88” is indicated, but this varies depend
-
ing on the location and environment of use and how long used.
Daily inspection
Perform daily inspection for use while paying attention to the following points.
Part to inspect
Inspection item
Check method
Mattress
Is sinking excessive?
* Some air cells will sink in the alternating pressure mode.
* If the mattress sinking is excessive, air cells may be damaged. Check the internal air cells, see
“
6
TROUBLESHOOTING” on page 21, and then take the necessary action.
Touch
inspection
Operation panel
(accessory)
• Is the power lamp lit in green?
• Is an error code displayed?
* If the power lamp is not lit in green or an error code is displayed, see “
6
TROUBLESHOOTING”
on page 21, and then take the necessary action.
Visual
inspection
Couplings
(upper and lower)
Are the couplings secured to the pump properly?
* If they are not secured properly, see “3. Connecting the pump and mattress” on page 6, and
then reconnect them.
Cable connectors
(pump)
• Are the cable and AC adapter connector connected properly?
* If they are not connected properly, see “2. Installing the pump and connecting a separately sold
operation panel (accessory)” on page 5, and then reconnect them.
• Are the caps attached properly (except when connecting the operation panel and bed link cable)?
* If they are detached, moisture and dust may enter inside and cause an electric shock or failure.
Attach them properly.
Regular inspection
To ensure safe use, inspect the product at least once a year. The product wears out and deteriorates differently depending on the
frequency of use and environment. Regularly inspect the following items. When an irregularity has been noticed or when a detailed
inspection is desired, contact Paramount Bed (see page 31) or your distributor.
Part to inspect
Inspection item
Check method
Cover
Is there damage?
Visual
inspection
Inside mattress
• Are the snap buttons present?
• Are the blowing tubes and air cells connected?
• Are the couplings (upper and lower) and blowing tubes connected?
• Is the top cushion damaged?
Pump
• Is there damage?
• Is the filter dirty?
* If the filter is dirty, see “Replacing the filter” on page 19, and then replace the filter.
Couplings
(upper and lower)
Is there damage?
Cables
Are there scratches and damage?
Operation panel
(accessory)
• Are the cables tangled in the air mattress or bed?
• Are any of the plugs and cables damaged?
• Is the operation panel scratched or damaged?
* Request repair (or replacement) of electrical parts such as the operation panel if they are damaged.
Position sensor
• Are there scratches and damage?
• Is it detached from the cover?
• Are the left and right sensor cover attachment zippers closed?
• Is the sensor sheet bent?