10

2. Starting use

 

„

How to start

Disconnect the power plug of the bed, then connect the bed link cable to the bed. See the procedure on pages 6 to 8.

 

„

How to shutdown

Disconnect the power plug of the bed and turn the battery off, then disconnect the bed link cable from the bed. 

 

„

Time required to complete air inflation

When you insert the power plug into an outlet or connect bed link cable to bed, the air cells will be automatically inflated with air 

after a short while. The “Preparing” indication will be lit during inflation. When air inflation completes, the “Preparing” indication 

turns off. (Time required to complete air inflation: 10 minutes)

 

„

When automatic softness control

When the user is lying in the bed, a softness (mattress internal pressure) that matches the weight, body shape, and lying po-

sition will be set automatically. The set softness is displayed for a while and then disappears. It can be displayed for a while by 

pressing the [Automatic softness control] button.

WARNING

When using automatic softness control, check that the width indicated on the quality label and 

the width setting indicated on the operation panel are the same (see page 11). If they differ, 

the body pressure may not be dispersed effectively.

Note

!

 

z

If the user leaves the bed when using automatic softness control, the softness may be automatically set to 

5. This is not a failure.

 

z

When the user is lying in bed, a softness that matches the user’s weight, body shape, and lying position will 

be set in about 5 to 15 minutes.

 

„

When manual softness control

Press the adjustment buttons to adjust the softness. To use automatic softness control, see page 11 and then switch to au-

tomatic softness control.

 

„

Fast inflation

I

nflate the mattress at high speed when power is supplied or max. inflate mode is set to ON.

*  The pump sound will be loud and the “Preparing” indication on the operation panel will be lit during fast inflation.

 

„

Functions retained after power-off

 

z

The function settings immediately before a power-off will be retained in the following cases.

•  When the power has recovered after a power failure

•  When restarted by disconnecting and reinserting the power plug or the bed link cable

 

z

The remaining time immediately before a power-off is retained for the max. inflate mode.

 

z

If “18^” was used with manual softness control, the softness setting becomes “18.”

 

„

Power failure countermeasure function

Automatically stop air discharge in the event of a power failure. Air can be held for two weeks, but this period varies depending on the 

conditions of use. However, the depressurization effect, alternating pressure function, etc. for the product do not work during a power 

failure, so be careful about the user’s condition. The small flow air cells (SF cells) maintain their expanded state. Contract them by 

disconnecting the coupling (upper) if necessary.

4

 FUNCTIONS AND OPERATING PROCEDURES

3. Setting of each function

3-1.  How to switch to the setting display

1

 Press the Setting button on the back of the operation panel.

2

 When you press the Setting button, the various setting buttons light and operation becomes possible. Unselected setting but-

tons of the various setting buttons are dimmed.

3

 

Press the Setting button on the back of the operation panel. The changes to the settings complete and the various setting buttons turn off.

Note

!

Even if you select various settings during setting display, changing of the settings will not complete if you do 

not press the Setting button on the back of the operation panel again. Check that the changes are reflected in 

the status display screen after changing settings.

1

 Press  the 

Setting button 

to switch 

to setting 

display.

2

 Press  the 

button of the 

setting you 

wish to set.

3

 Press  the 

Setting 

button to 

complete 

the setting.

Summary of Contents for cococia Riraku flow

Page 1: ...ease contact Paramount Bed or your distributor Intended Use cococia Riraku flow is an air mattress with an internal pump which is designed to disperse the body pressure to prevent pressure ulcers in m...

Page 2: ...attress and pump and do not allow a person s weight to be applied through their hips or knees The coupling may become disconnected resulting in air leak age and the body pressure not being dispersed e...

Page 3: ...mp have an expected service life of five years To maintain the condition of the pump have the pump serviced regularly according to the schedule recom mended Do not use unapproved accessories or attemp...

Page 4: ...nap button Product ID label affixed to side Position sensor cable Pump installation position Sheet stopper Coupling ports Operation panel The product code is printed on the back side Top Bottom Instru...

Page 5: ...d to the top bottom head end and foot end Failure to do so will result in not being able to disperse the body pressure effectively z Place the product directly on the bed and do not place anything oth...

Page 6: ...so Orient the plug so that the notch is that the notch is facing downward facing downward and insert it and insert it 3 3 Turn the connector cap Turn the connector cap clockwise clockwise 9 9 Turn the...

Page 7: ...zippers two locations and pull the operation panel cable and bed link cable out of the cover from the cor ner of the cover between the zippers WARNING z Check that the pump and couplings upper and low...

Page 8: ...e instruction manual of the bed to learn how to operate the bed and ensure you understand how to safely and properly use the bed before operating it Note 1Make the bottom of the bed flat horizontal 2R...

Page 9: ...e bed from the outlet and turn the battery off 3Orient the plug without warning label of the bed link cable with its protrusion facing up and insert it into the air mattress outlet of the bed 4Pass th...

Page 10: ...other than as sheet stoppers 5 Making the bed Towel mattress pad etc 4FUNCTIONS AND OPERATING PROCEDURES 1 Names and functions of operation panel parts 1Max inflate mode button page 13 2Power lamp Li...

Page 11: ...de is set to ON The pump sound will be loud and the Preparing indication on the operation panel will be lit during fast inflation Functions retained after power off z The function settings immediately...

Page 12: ...ying position and body shape in addition to the weight during automatic softness control Note z The table for Manual softness control shows the recommended softness level correspond ing to the recomme...

Page 13: ...SF cells z When back support is disabled all SF cells are disabled if the back angle becomes 45 or more SF sensor off mode If you insert the power plug into an outlet while pressing the Setting button...

Page 14: ...anual max inflate mode turns on the Max inflate mode button lights in blue and the status display screen displays the remaining time indication How to turn manual max inflate mode off Press the Max in...

Page 15: ...the type of bed and BED Navi used For details contact Paramount Bed or your distributor 1 button Displays the home screen 2AIR MATTRESS button Displays the AIR MATTRESS home screen see page 15 The fig...

Page 16: ...al softness control operation 4Preparing indication This is lit during internal pressure adjustment with fast infla tion 5SMALL FLOW SETTING button Changes the small flow setting 6SF cell status displ...

Page 17: ...F cell adjustment button to enable or disable the SF cell GO BACK RETURN button Confirms the settings and closes the INFLATE SETTING screen SF cell adjustment buttons Touch each SF cell adjustment but...

Page 18: ...he AIR MATTRESS home screen while automatic soft ness control is set 8MANUAL button Sets the manual softness control function is displayed in the AIR MATTRESS home screen while manual softness control...

Page 19: ...mold generation z If blood urine sebum and other stains are present use neutral detergent or soap to clean them and do not leave them If such stains are left as is the cover may deteriorate or develop...

Page 20: ...s attached or no filter is attached insects or dust may enter the inside of the pump possibly causing malfunction This also may cause failure Take out Pump side Filter Filter case Press hook How to cl...

Page 21: ...uplings secured to the pump properly If they are not secured properly see 3 Connecting the pump and mattress on page 6 and then reconnect them Cable connectors pump Are the cable and AC adapter connec...

Page 22: ...hile checking that there is no con tact with the mattress base 10 Is the user of the product a patient with a prominent bone projection hump back or lower limb contracture If a user with a prominent b...

Page 23: ...the length of the mattress Have the user sit down on the side of the bed in a position inside of the sensor detection range 13 Is the back section of the bed raised When the back angle of the bed is 2...

Page 24: ...it with the CPR zipper side facing the underside of the pump side Pump cover Put into the widest part of the spacer pad Bed link cable Operation panel accessory separately sold Instruction manual Pla...

Page 25: ...roduct code RC 2030W Protection code IEC 60529 IPx4 Size cm Width x Length x Thickness 5 6 x 5 6 x 17 including hook Product weight kg 0 2 Cable length cm 200 Materials Case ASA resin Hook POM resin S...

Page 26: ...n China Bed link cable Symbol Description Do not pinch or pull the bed link cable Bed link compatible Refer to the instruction manual Mark for restriction of the use of hazardous substances in electri...

Page 27: ...max inflate mode Automatic softness control Small flow function Preparing SF cell status display Symbol Description This product is an accessory for medical devices and such accessory is required to b...

Page 28: ...f the bed Use of RF communication devices within a distance of 30 cm 12 in may cause a decline in the performance of the bed n The bed cannot be used adjacent to other devices or in a stacked state If...

Page 29: ...lity environment The customer or user of the bed must use this bed in an environment specified here Immunity test IEC 60601 1 2 Test level Compliance level Electrostatic discharge ESD IEC 61000 4 2 8...

Page 30: ...electromag netic field IEC 61000 4 3 3 V m 80 MHz to 2 7 GHz 80 amplitude modulation 1 kHz 3 V m 80 MHz to 2 7 GHz 80 amplitude modulation 1 kHz a Field strengths from fixed transmitters such as thos...

Page 31: ...61000 4 3 27 V m 385 MHz 27 V m 380 to 390 TETRA400 Pulse modu lation 18 Hz 1 8 0 3 28 V m 450 MHz 28 V m 430 to 470 GMRS460 FRS460 FM 5 kHz devi ation 1 kHz sine 2 0 3 9 V m 710 MHz 9 V m 704 to 787...

Page 32: ...performance of the product in stock for at least 6 years after the discontinuation of production 3 Questions regarding after sales service Contact Paramount Bed see below Disposal When disposing of t...

Reviews: