910-198E
8/19/03
Page 9-2
Model 8500
The hardware data information will appear on the Measure mode
display.
If the user presses directly on the probe tip with a nonmagnetic
object, such as plastic, the INT Index number should not change by
more than two counts. Large changes that occur may indicate that the
tip is damaged and unable to provide reliable signals to the gage. If
the problem continues, try changing the probe and probe cable. To
clear the hardware data from the screen, go back to the Diagnostics
Setup menu and turn Debug to OFF.
9.2
Battery
The gage operates for at least eight (8) hours between charges under
normal conditions. A battery percentage is continuously shown in the
lower right corner of the display to represent the remaining battery
life. When the battery charge is insufficient, the gage automatically
turns-off to prevent damage to the battery. Recharge the battery using
the charger and line cord provided with the unit.