8/19/03
910-198E
Model 8500
Appendix V-3
As the thickness increases however, the following resolution error
values will supersede the fundamental error if greater than the
fundamental:
At 0.1200 inches or above, 0.0005
At 0.2000 inches or above, 0.0010
At 0.3000 inches or above, 0.0020
For example, at a thickness or 0.1600, the Basic Accuracy is + or -
(0.0048 + 0.0005) or 0.0053. A 0.1600" reference could, therefore,
read as low as 0.1547 and as high as 0.1653.
Please note that the Basic Accuracy is only guaranteed in a probe that
is in good or new condition. Tip distortion due to abrasive materials
or magnetic shock may cause the probe, during its lifetime, to not
meet Basic Accuracy specifications. Unless this distortion becomes
extreme however, inaccuracies can be corrected and Multi-Point
Accuracy can still be achieved by employing the Multi-Point
Calibration procedure described in Section 2 of this manual.