910-198E
8/19/03
Appendix V-2
Model 8500
•
The -105 probe has a tip modified to fit into the groove of an
automotive air-bag tear seam or similarly tight areas. The tar-
gets for these probes are 1/16" (1.6mm) diameter ball, a 3/16"
(4.8mm) diameter disk with a sharp edge, and a 1/2" (12.7mm)
diameter disk with a square 1/32" (0.79mm) wide edge. The
disks are used exactly the same way as target balls, but align
with the shape of the tip.
•
The -109 probe has a 1/4" (6.4mm) radius tip, and is used for
measurements up to 0.400" (10mm). The targets employed with
this probe are a 1/8" (3.2mm) diameter ball, a 3/16" (4.8mm)
diameter ball and a 1/4" (6.4mm) diameter ball. The -109 is a
type 6 probe.
•
The -110 probe is the same as -109, except that the connector is
mounted on the side, 90-degrees from its normal location.
•
The -111 probe is a standard probe with a super-wear resistant
material added to the tip. This probe uses standard targets.
•
The -112 probe is the same as -111, except that the connector is
mounted on the side, 90-degrees from its normal location.
The 80FXV, V-standard plastic fixture, fits over any probe tip, and
helps to align some items over the probe axis, such as tubing.
The accuracy specification for the -109 and -110 extended range probes
should be quoted for all target balls (1/8”, 3/16”, and 1/4”) as follows:
Basic Accuracy: + or -3%, + or - ( a resolution error)
Reference Accuracy: + or - 1%, + or - ( a resolution error)
The “resolution error” is described as the greater of the following values:
Fundamental resolution error:
High resolution mode: 0.0001; Low resolution mode: 0.001
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com