910-198E
8/19/03
Page 5-14
Model 8500
The two button probe is also an optional item. By default, the Max
button on a two-button probe is programmed to function as the send
command; however, this functionality may be altered to the Meas or
Save function. To view or change the Max button configuration, press
[SETUP] and go to the Communications menu.Use the [ ] and [ ]
keys to highlight Max Key and the [
] or [
] key to choose either
Save, Send, or Meas.
The Save/Send function in the Communication Setup menu can be set
to Primary or Secondary. For more information on these parameters,
see Section 4.1, Saving and Sending Values.
5.4
Calibration Setup
The Calibration Setup menu contains the calibration setup parameters
for the gage. To enter the Setup menu, press [SETUP]. Use the [ ] and
[ ] keys to select Calibration and press [ENTER]. The gage displays
the Calibration Setup menu:
Use the [ ] and [ ] keys to highlight the parameter and the [
] or
[
] key to select a setting:
Auto QCAL - The Auto QCAL function is responsible for the
automatic, internal calibration that compensates for probe temperature
fluctuations. When this function is turned ON, it will allow the gage to