USER'S MANUAL
General instructions for use
I-Max Touch (110-120V)
(Rev. 0)
64
6.6.3.2 Carrying out the first exposure (mouth closed)
WARNING:
During the emission of X-rays, the protection procedures for the operator
and personnel in the area must be in compliance with the local
regulations.
In all cases, it is recommended that during the emission of X-rays, only
the patient and operator be present in the room. If the operator is not
protected by suitable screens, he must stand at least 1.5 m (4.9 ft) away
from the emission of the rays (see the Figure 1 and Figure 2).
NOTE:
If deemed necessary, it is possible to check the interference of the
rotation movement with the shoulder of the patient; it is possible,
by pressing the key "Test" (5) , to activate the Test function.
In this condition, it will be possible to make the machine perform all the
movements made during the examination, but
without emitting rays.
The test function of the TMJ closed/open mouth is the same as for the
panoramic mode and so there will not be a second rotation corresponding
to the open mouth exam. Once the cycle is completed, deactivate the
"Test" function by pressing the key again.
1.
Check once again that the exposure data are correct. If not, correct
them as described in paragraph 6.6.2.2. ensure that the machine's
indicator light "Ready for X-ray" will come on, so press the X-ray
button for the entire duration of the exposure, checking the
simultaneous working of the ray indicator light "X-ray emission" (if
you are within sight of the machine) and the acoustic ray signal. The
following message will be displayed first:
S T A R T E X A M
P R E - H E A T I N G . . .
and then (after 2 seconds), the following message will be displayed:
x x k V x x m A 9 . 7 0 s
>
X
- R
A
Y
<
x = value defined by the settings
NOTE:
If the machine is in the "Test" mode, the display will show:
T E S T
X R A Y N O T A C T I V E
*
*
Summary of Contents for I-Max Touch