USER'S MANUAL
Technical characteristics
I-Max Touch (110-120V)
(Rev. 0)
14
5. TECHNICAL
CHARACTERISTICS
General features
Type I-Max
Touch
Manufacturer OWANDY
Champs sur Marne, France
Class
Class II B for European Directive
for Medical Devices 93/42
Class II for Canadian MDR
Class I with type B applied parts
according to IEC 60601-1
Class II according to 21CFR-
subchapter J
Protection degree
IPX0 standard device
Rated line voltage
110-120 V
∼
Line frequency
50/60 Hz
Maximum line current
15 A @ 115 V
∼
50/60 Hz
Power consumption
1.7 kVA @ 115 V
∼
50/60 Hz
Protection fuse (F1)
15 A T
Protection fuse (F2)
of switching power
supply
3 A T
HF generator board protection fuses
F1: 10 A F
F2: 5 A HF
F3: 2 A T
Line voltage regulation
< 3 % at 99 V
∼
Rated output voltage (kVp)
60 ÷ 86 kV
p,
with 2 kV
p
steps
Anodic current
6 ÷ 10 mA, with 1 mA steps for PAN, TMJ,
and Sinus
6 ÷ 12 mA in 1 mA steps for Ceph (up to
76 kVp)
6 ÷ 10 mA in 1 mA steps for Ceph (from
78 kVp to 86 kVp)
Sensor cover additional filtration
0.5 mm Al eq @ 70 kV
p
Summary of Contents for I-Max Touch
Page 1: ...Version 25 September 2009 Rev 0 I Max Touch 0459 110 120V version User s Manual...
Page 5: ...USER S MANUAL Contents I Max Touch 110 120V Rev 0 iv THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 129: ...USER S MANUAL Maintenance I Max Touch 110 120V Rev 0 124 THIS PAGE IS INTENTIONALLY LEFT BLANK...