USER'S MANUAL
General instructions for use
I-Max Touch (110-120V)
(Rev. 0)
68
6.6.4.2 Carrying out the second exposure (mouth open)
WARNING:
During the emission of X-rays, the protection procedures for the operator
and personnel in the area must be in compliance with the local
regulations.
In all cases, it is recommended that during the emission of X-rays, only
the patient and operator be present in the room. If the operator is not
protected by suitable screens, he must stand at least 1.5 m (4.9 ft) away
from the emission of the rays (see the Figure 1 and Figure 2).
WARNING:
Using the laser centring devices, check that the system is still aligned
with the patient's sagittal medial plane.
1.
Press the key "Patient Entrance" (6) . The display will show:
x x k V x x m A 9 . 7 0 s
S T A R T E X A M
Check again that the exposure data are correct (see paragraph 6.6.2).
NOTE:
The Adult/Child and Size
small - medium - large
selection keys are
deactivated. The exposure parameters can be changed as described in
paragraph 6.6.2.
Press the X-ray button for the entire duration of the exposure,
checking the concurrent working of the ray indicator light "X-ray
emission" (if you are within sight of the machine) and the acoustic
ray signal. The following message will be displayed first:
S T A R T E X A M
P R E - H E A T I N G . . .
and then (after 2 seconds), the following message will be
displayed:
x x k V x x m A 9 . 7 0 s
>
X
- R
A
Y
<
x = value defined by the settings
*