USER'S MANUAL
General instructions for use
(Rev. 0)
I-Max Touch (110-120V)
43
6.5 Panoramic
examination
When making a panoramic examination, the tube-head support arm
(X-rays generator) makes a continuously rotating movement.
After an initial acceleration, rotation is made at a constant speed. At the
same time the arm carries out a scanning towards the support column.
The X-rays are emitted only when the rotating arm is at constant rotation
speed.
The patient's centring is assisted by two linear luminous laser beams,
which indicate the position of the sagittal medial plane; the
corresponding patient's plane needs to be aligned with this plane. The
latter is held in place, during the examination phase, by means of temple
clasps rods and of a forehead support. A fourth fixing point is determined
by the chin support.