6.5.2.1 Anatomic
exposure
Select the type of patient with the
Adult/Child
key (7).
Select the type of build with the
Size
(8) key (
small - medium - large).
On the basis of these selections, the display will visualise the kV and mA
settings as in the table.
Exposure values in PAN mode
Adult Patient
(13.8 seconds)
Child Patient
(13.4 seconds)
kV mA kV mA
Small
68 6 64 6
Medium
72 6 66 6
Large
74 6 68 6
Table 2
Select the type of biting with the key "Type of Biting Selection"
(9) .
NOTE:
The type of biting does not affect the kV and mA values, but it affects the
position of the focus layer, by adapting the rotation movement to the
patient's anatomy.
*