USER'S MANUAL
General instructions for use
I-Max Touch (110-120V)
(Rev. 0)
60
6.6.2.1 Anatomic
exposure
Select the type of patient with the
Adult/Child
key (7).
Select the type of build with the
Size
(8) key (
small - medium - large).
On the basis of the selections made, the display will visualise the kV and
mA settings as in the table.
Exposure values in TMJ examination (9.7 sec)
Examination
Adult Child
TMJ mouth closed/open
kV mA kV mA
Small
68 6 62 6
Medium
72 6 64 6
Large
76 6 66 6
Table 3
The time (9.7 sec.) refers to the sum of the four exposures (2 closed
mouth exposures and 2 open mouth exposures).
Summary of Contents for I-Max Touch
Page 1: ...Version 25 September 2009 Rev 0 I Max Touch 0459 110 120V version User s Manual...
Page 5: ...USER S MANUAL Contents I Max Touch 110 120V Rev 0 iv THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 129: ...USER S MANUAL Maintenance I Max Touch 110 120V Rev 0 124 THIS PAGE IS INTENTIONALLY LEFT BLANK...