USER'S MANUAL
General instructions for use
(Rev. 0)
I-Max Touch (110-120V)
31
Furthermore there is the possibility to
manually select the exposure
parameters; in this case, it is possible
to set the parameter with the desired
value.
The parameters available are: kV and
mA (Soft Tissue Filter position, mm,
only in cephalometry).
When the exposure parameters are
changed manually, the mode indicator
switches from "Anatomic" to "Manual".
Return to "Anatomic mode" using the
main programme selection button. By
holding the indicator for more than
1 second the special mode is activated
and the indicator changes colour. In
special mode modified exposure
parameters can be stored or exposure
verification tests can be performed.
There are two light indicators; the first
one on the top indicates the condition
"Machine Ready", indicating the user
that by pressing the X-ray button key
once more, X-rays emission will start;
the second indicates the effective
emission of X-rays.
The movement of the column is
controlled by the appropriate keys.
The speed has two set values.
The movements are enabled during
equipment setting.
The key "Luminous centring device"
helps turn ON/OFF the laser centring
devices that allow the correct
positioning of the medial-sagittal and
Frankfurt planes, by adapting the
I-Max Touch to the patient's anatomy.
Test OFF
Test ON
The key "Test" is used to avoid the
X-rays emission, in order to check the
absence of collisions with the patient.