USER'S MANUAL
Technical characteristics
(Rev. 0)
I-Max Touch (110-120V)
27
3.
The KV and mA parameters can be modified by pressing the increase
key and the decrease key of the KV and mA on the display.
The parameters can vary within the limits shown in the following
table:
Parameter
Minimum value
Maximum value
kV
60 86
mA
6 12
Table 1
4.
Perform an exposure; technical factors can be read on the measuring
instrument.
NOTE:
The performance is guaranteed only if the measurement of kV and time is
done with the invasive method, due to the fact that the non-invasive
method may introduce errors for instruments tolerance or wrong
measurement condition.
5.
To end the control program, press any key other than the increase
and decrease keys ; the display will indicate:
x x k V x x m A x x . x s
P A N O R A M I C - S T D
and the unit will return to standard mode.
*
Summary of Contents for I-Max Touch
Page 1: ...Version 25 September 2009 Rev 0 I Max Touch 0459 110 120V version User s Manual...
Page 5: ...USER S MANUAL Contents I Max Touch 110 120V Rev 0 iv THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 129: ...USER S MANUAL Maintenance I Max Touch 110 120V Rev 0 124 THIS PAGE IS INTENTIONALLY LEFT BLANK...