USER'S MANUAL
General instructions for use
I-Max Touch (110-120V)
(Rev. 0)
88
6.8.4
Making an exposure
WARNING:
During the emission of X-rays, the protection procedures for the operator
and personnel in the area must be in compliance with the regulations in
force in the country where the machine is used.
In all cases, it is recommended that during the emission of X-rays, only
the patient and operator be present in the room. If the operator is not
protected by suitable screens, he must stand at least 1.5 m (4.9 ft) away
from the emission of the rays (see Figure 1 and Figure 2).
1.
Verify once again that the exposure data are correct (see paragraph
6.8.2). Advise the patient to remain still and to keep his mouth
closed, with the teeth touching, throughout the duration of the
exposure.
Press the "Patient Entrance" (6) key .
The unit will move into the selected examination start position.
The signalling LED "Ready for X-ray" will light up, indicating that the
machine is ready to produce X-rays.
NOTE:
If you want to cancel the operation, press key "Patient Entrance" (6)
2.
Press the X-ray button for the entire duration of the exposure,
checking the concurrent working of the ray indicator light "X-ray
emission" (if you are within sight of the machine) and the acoustic
ray signal.
The following message will be displayed first:
S T A R T E X A M
P R E - H E A T I N G . . .
and then (after 2 seconds), the following message will be displayed:
x x k V x x m A x . x x s
>
X
- R
A
Y
<
x = value defined by the settings
*