User Manual
– Technical characteristics
NIMXEN020I
Owandy Radiology SAS
24
Exposure times
Panoramic exam (PAN)
14 s Adult / 12.8 s Child
Emi-panoramic exam
7.7 s Adult / 7.1 s Child
Improved orthogonality panoramic exam
11.5 s Adult / Child
Reduced dose panoramic exam
11.6 s Adult / 10.4 s Child
Frontal dentition
4.1 s Adult / Child
Bitewing R, Bitewing L
3.1 s Adult / Child
Bitewing R&L
6.2 s Adult / Child
TMJ mouth closed/open
10.6 s for left and right joint in open and closed
condition
TMJ single phase
5.3 s
Sinus P/A projection
9 s
3D exams (except TMJ 3D)
5.7 s
TMJ 3D
5.0 s
Exposure time accuracy
± 5 % or ± 20ms whichever is greater
Exam modes
Exam selection
•
Automatic selection for Adult and Child, 3
Sizes
•
3 biting modes (Panoramic exam)
•
Manual selection
Panoramic exam
NOTE:
Some of these exams are optional and
depend on the system configuration.
•
Standard panoramic
•
Half panoramic L/R
•
Ortho Rad panoramic
•
Low dose panoramic
•
Frontal dentition
•
Bitewing Left/Right
•
Bitewing Left and Right
TMJ (Temporal Mandibular Joint) exam
TMJ closed and open mouth
Sinus exam
Sinus P/A projection
Volumetric 3D exams
Automatic selection for Adult and Child, 3 sizes
chosen between: 3D Full Dentition, 3D Single
Jaw
(Mandibular/Maxillary),
3D
Mandibular/Maxillary Teeth (frontal, premolar,
molar), 3D TMJ (Left/Right), 3D Sinus
3D Dentition reconstructed volume
3D Full Dentition (
*
)
85 mm x 93 mm (Diameter x Height)
3D Single Jaw (Mandibular and Maxillary) (
*
)
85 mm x 50 mm (Diameter x Height)
3D Mandibular/Maxillary Teeth
50 mm x 50 mm (Diameter x Height)
3D Extended volume
116 mm x 103 mm (Diameter x Height)
(*) In case the 80x80 limitation is set, the values will change to: 3D Full Dentition 80 mm x 80 mm
(Diameter x Height); 3D Single Jaw (Mandibular and Maxillary) 80 mm x 50 mm (Diameter x
Height)