User Manual
– Making an exam
NIMXEN020I
Owandy Radiology SAS
74
10.1.1
Anatomic exposure
Select the type of patient with the Adult/Child icons. If child option is selected, exposure
parameters are lower respect to the corresponding adult programs. In addition, the exam
trajectory in panoramic programs is reduced of about 10%.
Select the type of build with the Size icons (small - medium - large).
On the basis of these selections, the display will show the kV and mA settings accordingly.
Select the type of biting with the icon "Type of Biting Selection" (option available in Panoramic
mode only).
Default
exposure values in 2D Panoramic modes
Adult Patient
(14 seconds)
Child Patient
(12.8 seconds)
kV
mA
kV
mA
Small
76
9
66
8
Medium
80
9
68
8
Large
82
9
70
8
Default
exposure values in 2D Sinus mode
Adult Patient
(9 seconds)
Child Patient
(9 seconds
)
kV
mA
kV
mA
Small
68
8
64
8
Medium
72
8
66
8
Large
74
8
68
8
Default
exposure values in 2D TMJ mode
Adult Patient
(10,6 seconds)
Child Patient
(10,6 seconds
)
kV
mA
kV
mA
Small
70
8
64
8
Medium
74
8
66
8
Large
78
8
68
8
Exposure values in 3D Full Dentition, 3D Extended Volume and 3D Airways
Adult Patient
(7 seconds)
Child Patient
(7 seconds
)
kV
mA
kV
mA
Small
84
4
64
6.3
Medium
84
5
66
6.3
Large
84
6
68
6.3