– Quality assurance program
NIMXEN020I
Owandy Radiology SAS
52
4.
Place the support plate on the chin rest and place the 3D quality phantom on the plate, in
such a way as the reference central line is on the top.
3D Quality phantom
Support plate
Figure 13
5.
On the PVC insertion is present a position reference; this reference must be positioned
towards the keyboard side.
Figure 14
Position
reference