Select "2D" exam.
8.
Make an exposure at 66 kV, 6.3 mA (see chapter 10).
9.
Select the "Ruler" icon and measure the distance between the two external spheres; this
value must be 173mm ± 2mm.
In case the test fails, call technical assistance.
10. Record the tests results in the log book at paragraph 8.4.1.